» Articles » PMID: 11271080

Different Dosages of Dobutamine in Septic Shock Patients: Determining Oxygen Consumption with a Metabolic Monitor Integrated in a Ventilator

Overview
Specialty Critical Care
Date 2001 Mar 29
PMID 11271080
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: Oxygen consumption (VO2) obtained from respiratory gases by indirect calorimetry (VO2,IC) with a metabolic monitor integrated in a ventilator were to be compared to VO2 obtained by the Fick principle (VO2,Fick) in septic patients following an increase in oxygen delivery (DO2) induced by positive inotropic support.

Design: Prospective clinical study.

Setting: University Hospital, Surgical Intensive Care Unit (ICU).

Patients: Thirty patients suffering from sepsis.

Interventions: DO2 was increased by dobutamine infusion, starting with an initial dosage of 5 microg x kg x min, increased to a maximum of 10 microg x kg x min.

Measurements And Main Results: Dobutamine infusion induced a dosage-related increase in DO2 (from 577 +/- 192 to 752 +/- 202 ml x min x m2, p < 0.01), which was associated with a statistically significant increase in VO2,IC (from 173 +/- 30 to 188 +/- 28 ml x min x m2, p < 0.01) and in VO2,Fick (from 140 +/- 25 to 156 +/- 24 ml x min x m2, p < 0.01). The comparison between VO2,IC and VO2,Fick revealed differences (bias and precision--33 +/- 32 ml x min x m2).

Conclusions: With a metabolic monitor integrated in a ventilator it was possible to carry out continuous monitoring of calorimetric data under clinical conditions. In contrast to previous studies using indirect calorimetry, this study showed a moderate correlation between VO2 and DO2 in septic patients using either method. The clinical relevance of this finding requires further investigation. Different factors (e. g. injectant temperature, pulmonary VO2) produced substantial differences between VO2,IC and VO2,Fick as previously shown.

Citing Articles

Perioperative estimations of oxygen consumption from LiDCO™plus-derived cardiac output and Ca-cvO2 difference: Relationship with measurements by indirect calorimetry in elderly patients undergoing major abdominal surgery.

Jakobsson J, Noren C, Hagel E, Backheden M, Kalman S, Bartha E PLoS One. 2024; 19(7):e0272239.

PMID: 39052563 PMC: 11271938. DOI: 10.1371/journal.pone.0272239.


The Effects of Dobutamine in Septic Shock: An Updated Narrative Review of Clinical and Experimental Studies.

Dubin A, Mugno M Medicina (Kaunas). 2024; 60(5).

PMID: 38792934 PMC: 11123338. DOI: 10.3390/medicina60050751.


Hemodynamic effects and tolerance of dobutamine for myocardial dysfunction during septic shock: An observational multicenter prospective echocardiographic study.

Razazi K, Labbe V, Laine L, Bedet A, Carteaux G, de Prost N Front Cardiovasc Med. 2022; 9:951016.

PMID: 36158835 PMC: 9500364. DOI: 10.3389/fcvm.2022.951016.


Methodological Aspects of Indirect Calorimetry in Patients with Sepsis-Possibilities and Limitations.

Wasyluk W, Zwolak A, Jonckheer J, De Waele E, Dabrowski W Nutrients. 2022; 14(5).

PMID: 35267903 PMC: 8912694. DOI: 10.3390/nu14050930.


In-hospital mortality following treatment with red blood cell transfusion or inotropic therapy during early goal-directed therapy for septic shock: a retrospective propensity-adjusted analysis.

Mark D, Morehouse J, Hung Y, Kene M, Elms A, Liu V Crit Care. 2014; 18(5):496.

PMID: 25212411 PMC: 4181703. DOI: 10.1186/s13054-014-0496-y.