» Articles » PMID: 11270701

Extension of Generator Longevity by Use of High Impedance Ventricular Leads

Overview
Date 2001 Mar 29
PMID 11270701
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

The resistance of a pacing lead negatively correlates to current consumption. A prospective, randomized trial was conducted to evaluate the effect of a high impedance ventricular lead (CapSure Z) on generator longevity compared to a conventional lead (CapSure SP) eighty-nine patients were included in the study (51 male, 37 female, age 70.0+/-10.3 years). Forty-six patients received a CapSure SP lead (5024 bipolar), and 43 patients received a CapSure Z lead (5034 bipolar) in a randomized fashion. Follow-up data collected at 5 days, 3, 6, and 12 months postimplant included: lead impedance, pacing and sensing thresholds, impulse energy, and estimated time to replacement. All parameters were collected via pacemaker telemetry; the time to replacement was calculated automatically by a programmed algorithm of the pacemaker. There was no difference in the performance of the atrial lead when a dual chamber device was indicated. The CapSure Z leads displayed statistically significant higher impedance values than the CapSure SP lead in all follow-up periods. There was no significant difference in lead related complications. No significant differences were observed between pacing and sensing thresholds in both groups. The CapSure Z leads provided a significant reduction in current drain, resulting in a reduction of mean energy consumption at the 12-month follow-up from 10.4+/-5.0 microJ in the CapSure SP group to 6.6+/-1.4 microJ in the CapSure Z group (median from 9.9 microJ to 6.9 microJ, respectively), providing an estimated increase in mean longevity of more than 1 year from 81.1+/-23.5 months in the CapSure SP group to 94.5+/-13.4 months in the CapSure Z group (median: 76.5 months to 95.0 months, respectively). The use of a high resistance lead for ventricular pacing appears to result in a clinically relevant extension of generator longevity.

Citing Articles

Energy Saving in Permanent Cardiac Pacing: Pulse Waveform and Charge Balancing Deserve Consideration.

Di Gregorio F, Marcantoni L, Mozzi A, Barbetta A, Zanon F Bioengineering (Basel). 2025; 12(2).

PMID: 40001713 PMC: 11851791. DOI: 10.3390/bioengineering12020194.


Combining current of injury and P-wave sensing optimized right atrial active-fixation leads implantation.

Chen M, Zhang F, Wang W, Chen X, Chen J, Chen L J Thorac Dis. 2019; 11(4):1279-1286.

PMID: 31179070 PMC: 6531755. DOI: 10.21037/jtd.2019.04.04.


Pacemaker longevity: the world's longest-lasting VVI pacemaker.

Katz D, Akiyama T Ann Noninvasive Electrocardiol. 2007; 12(3):223-6.

PMID: 17617067 PMC: 6931941. DOI: 10.1111/j.1542-474X.2007.00165.x.