Accuracy of Plastic Replica of Aortic Aneurysm Using 3D-CT Data for Transluminal Stent-grafting: Experimental and Clinical Evaluation
Overview
Authors
Affiliations
Purpose: To plan stent-grafting for aortic aneurysms with complicated morphology, we prepared life-sized aortic replicas by laser stereolithography using helical 3D--CT data. The accuracy of the replica was evaluated by measurement of vessel phantoms and clinical 3D--CT data.
Method: An imaginary aortic wall was created from helical CT images of the aorta, and a hollow plastic replica was produced by laser stereolithography. The accuracy of the replica was evaluated in five abdominal aortic aneurysms by experimental phantom studies and measurements of the replicas.
Results: The mean difference in measurements between 3D--CT images and model vessels and between 3D--CT images and aortic replicas was 0.2 mm each. Therefore, the difference in measurements between real aortic aneurysms and the replicas was at most 0.4 mm.
Conclusion: The accuracy of the replica is satisfactory, making it useful for preoperative evaluation and simulation for stent-grafting.
Rynio P, Wojtun M, Wojcik L, Kawa M, Falkowski A, Gutowski P Quant Imaging Med Surg. 2022; 12(2):1385-1396.
PMID: 35111632 PMC: 8739108. DOI: 10.21037/qims-21-529.
3D Bioprinting of cardiac tissue and cardiac stem cell therapy.
Alonzo M, AnilKumar S, Roman B, Tasnim N, Joddar B Transl Res. 2019; 211:64-83.
PMID: 31078513 PMC: 6702075. DOI: 10.1016/j.trsl.2019.04.004.
Cardiac 3D Printing and its Future Directions.
Vukicevic M, Mosadegh B, Min J, Little S JACC Cardiovasc Imaging. 2017; 10(2):171-184.
PMID: 28183437 PMC: 5664227. DOI: 10.1016/j.jcmg.2016.12.001.
Markl M, Schumacher R, Kuffer J, Bley T, Hennig J MAGMA. 2005; 18(6):288-92.
PMID: 16369802 DOI: 10.1007/s10334-005-0019-6.
Advanced "tactile" medical imaging for separation surgeries of conjoined twins.
Christensen A, Humphries S, Goh K, Swift D Childs Nerv Syst. 2004; 20(8-9):547-53.
PMID: 15278381 DOI: 10.1007/s00381-004-0982-7.