Imaging Characteristics of an Amorphous Silicon Flat-panel Detector for Digital Chest Radiography

Overview
Journal Radiology
Specialty Radiology
Date 2001 Mar 7
PMID 11230640
Citations 17
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate the imaging characteristics of an amorphous silicon flat-panel detector (FPD) for digital chest radiography.

Materials And Methods: The 41 x 41-cm digital FPD is constructed on a single monolithic glass substrate with a structured cesium iodide scintillator layer and an amorphous silicon thin-film transistor array for image readout. Basic imaging characteristics of the FPD and associated image processing system were assessed on acquired images, including linearity, repeatability, uniformity of response, modulation transfer function (MTF), noise power spectrum, detective quantum efficiency (DQE), contrast sensitivity, and scatter content. Results with the FPD system were compared to those with a storage phosphor computed radiography (CR) system.

Results: Images obtained with the FPD demonstrated excellent uniformity, repeatability, and linearity, as well as MTF and DQE that were superior to those with the storage phosphor CR system. The contrast and scatter content of images acquired with the FPD were equivalent to those acquired with the storage phosphor system.

Conclusion: The FPD provides radiographic images with excellent inherent physical image quality.

Citing Articles

Development of a Dedicated X-Ray Fluoroscopic Apparatus for Therapeutic Pancreatobiliary Endoscopy: A Review.

Fukuma T, Ishii S, Fujisawa T, Takahashi K, Nakamura T, Shibata F J Clin Med. 2025; 14(4).

PMID: 40004745 PMC: 11856850. DOI: 10.3390/jcm14041214.


Detectability of simulated interstitial pneumonia on chest radiographs: comparison between irradiation side sampling indirect flat-panel detector and computed radiography.

Yano Y, Yabuuchi H, Kairada A, Tanaka N, Morishita J, Akasaka T Br J Radiol. 2014; 87(1040):20140075.

PMID: 24874767 PMC: 4112402. DOI: 10.1259/bjr.20140075.


Estimation of the two-dimensional presampled modulation transfer function of digital radiography devices using one-dimensional test objects.

Wells J, Dobbins J Med Phys. 2012; 39(10):6148-60.

PMID: 23039654 PMC: 3465356. DOI: 10.1118/1.4752442.


Comparison of scatter rejection and low-contrast performance of scan equalization digital radiography (SEDR), slot-scan digital radiography, and full-field digital radiography systems for chest phantom imaging.

Liu X, Shaw C, Lai C, Wang T Med Phys. 2011; 38(1):23-33.

PMID: 21361171 PMC: 3017579. DOI: 10.1118/1.3519903.


Dynamic dual-energy chest radiography: a potential tool for lung tissue motion monitoring and kinetic study.

Xu T, Ducote J, Wong J, Molloi S Phys Med Biol. 2011; 56(4):1191-205.

PMID: 21285477 PMC: 3164987. DOI: 10.1088/0031-9155/56/4/019.