A Frictional Study of Total Hip Joint Replacements

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2000 Dec 29
PMID 11131195
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Polymeric wear debris produced by articulation of the femoral head against the ultra-high-molecular-weight polyethylene socket of a total hip replacement has been implicated as the main cause of osteolysis and subsequent failure of these implants. Potential solutions to this problem are to employ hard bearing surface combinations such as metal-on-metal or ceramic-on-ceramic prostheses. The aim of this study was to investigate the difference in lubrication modes and friction of a range of material combinations using synthetic and biological fluids as the lubricants. The experimental results were compared with theoretical predictions of film thicknesses and lubrication modes. A strong correlation was observed between experiment and theory when employing carboxy methyl cellulose (CMC) fluids as the lubricant. Under these conditions the ceramic-on-ceramic joints showed full fluid film lubrication while the metal-on-metal, metal-on-plastic, diamond-like carbon-coated stainless steel (DLC)-on-plastic and ceramic-on-plastic prostheses operated under a mixed lubrication regime. With bovine serum as the lubricant in the all ceramic joints, however, the full fluid film lubrication was inhibited due to adsorbed proteins. In the metal-on-metal joints this adsorbed protein layer acted to reduce the friction while in the ceramic coupling the friction was increased. The use of bovine serum as the lubricant also significantly increased the friction in both the metal-on-plastic and ceramic-on-plastic joints. The friction produced by the DLC-on-plastic joints depended on the quality of the coating. Those joints with a less consistent coating and therefore a higher surface roughness gave significantly higher friction than the smoother, more consistently coated heads.

Citing Articles

Tribological Characteristics of a-C:H:Si and a-C:H:SiO Coatings Tested in Simulated Body Fluid and Protein Environment.

Jedrzejczak A, Szymanski W, Kolodziejczyk L, Sobczyk-Guzenda A, Kaczorowski W, Grabarczyk J Materials (Basel). 2022; 15(6).

PMID: 35329533 PMC: 8951404. DOI: 10.3390/ma15062082.


Hip prosthetic loosening: A very personal review.

Mjoberg B World J Orthop. 2021; 12(9):629-639.

PMID: 34631447 PMC: 8472441. DOI: 10.5312/wjo.v12.i9.629.


Amorphous Carbon Coatings for Total Knee Replacements-Part II: Tribological Behavior.

Rothammer B, Marian M, Neusser K, Bartz M, Bohm T, Krauss S Polymers (Basel). 2021; 13(11).

PMID: 34198895 PMC: 8201056. DOI: 10.3390/polym13111880.


Postoperative changes in in vivo measured friction in total hip joint prosthesis during walking.

Damm P, Bender A, Bergmann G PLoS One. 2015; 10(3):e0120438.

PMID: 25806805 PMC: 4373913. DOI: 10.1371/journal.pone.0120438.


Friction in total hip joint prosthesis measured in vivo during walking.

Damm P, Dymke J, Ackermann R, Bender A, Graichen F, Halder A PLoS One. 2013; 8(11):e78373.

PMID: 24260114 PMC: 3832636. DOI: 10.1371/journal.pone.0078373.