Cardiac Output Measurement by Thermal Dilution: Reproducibility and Comparison with the Dye-dilution Technique

Overview
Journal Ann Surg
Specialty General Surgery
Date 1976 Jan 1
PMID 1108813
Citations 17
Authors
Affiliations
Soon will be listed here.
Abstract

Cardiac output estimates by the principle of thermodilution (COth) was compared with dye-dilution estimates (COdye) in pigs. For COth estimates a Swan-Ganz 7 F floating thermodilution catheter and a 9500 Edwards Computer, were used. The COdye estimates were obtained by the apparatus constructed by Zijlstra and Mook. The effect of the thermistor position in the pulmonary artery on the COth estimates was also investigated. The reproducibility of COth was examined by duplicate determinations. Based on 101 simultaneous estimates of COth and COdye the correlation was found COth = 1.020 COdye + 0.2378, r = 0.971 for cardiac outputs between 0.65 l/min and 11 l/min. For 111 duplicate determinations of COth between 2 and 9 l/min the coefficient of variation was 4.74%. The thermistor position in the pulmonary artery had no influence on the COth estimates provided an undamped pressure curve could be monitored from the tip of the catheter. Cardiac output can thus be measured rapidly with good accuracy also for low values by means of a blindly inserted thermistor catheter positioned without x-ray control and a computer with digital display.

Citing Articles

Continuous Maternal Hemodynamics Monitoring at Delivery Using a Novel, Noninvasive, Wireless,PPG-Based Sensor.

Atzmon Y, Ben Ishay E, Hallak M, Littman R, Eisenkraft A, Gabbay-Benziv R J Clin Med. 2020; 10(1).

PMID: 33375211 PMC: 7793094. DOI: 10.3390/jcm10010008.


Maternal cardiovascular hemodynamics in normotensive versus preeclamptic pregnancies: a prospective longitudinal study using a noninvasive cardiac system (NICaS™).

Lavie A, Ram M, Lev S, Blecher Y, Amikam U, Shulman Y BMC Pregnancy Childbirth. 2018; 18(1):229.

PMID: 29898711 PMC: 6001131. DOI: 10.1186/s12884-018-1861-7.


Cardiac hemodynamics before, during and after elective cesarean section under spinal anesthesia in low-risk women.

Ram M, Lavie A, Lev S, Blecher Y, Amikam U, Shulman Y J Perinatol. 2017; 37(7):793-799.

PMID: 28406485 DOI: 10.1038/jp.2017.53.


Pulse waveform analysis in the optic nerve head circulation reflects systemic vascular resistance obtained via a Swan-Ganz catheter.

Shiba T, Takahashi M, Hashimoto R, Matsumoto T, Hori Y Graefes Arch Clin Exp Ophthalmol. 2016; 254(6):1195-200.

PMID: 26887825 DOI: 10.1007/s00417-016-3289-y.


Comparison between cardiac output values measured by thermodilution and partial carbon dioxide rebreathing in patients with acute lung injury.

Valiatti J, Amaral J Sao Paulo Med J. 2005; 122(6):233-8.

PMID: 15692716 PMC: 11126179. DOI: 10.1590/s1516-31802004000600002.


References
1.
Rahimtoola S, Swan H . CALCULATION OF CARDIAC OUTPUT FROM INDICATOR-DILUTION CURVES IN THE PRESENCE OF MITRAL REGURGITATION. Circulation. 1965; 31:711-8. DOI: 10.1161/01.cir.31.5.711. View

2.
JOSE A, McGaff C, MILNOR W . The value of injections of dye into the left heart in the study of mitral and aortic valvular disease by catheterization of the left heart. Am Heart J. 1960; 60:408-16. DOI: 10.1016/0002-8703(60)90200-3. View

3.
EVONUK E, IMIG C, Greenfield W, Eckstein J . Cardiac output measured by thermal dilution of room temperature injectate. J Appl Physiol. 1961; 16:271-5. DOI: 10.1152/jappl.1961.16.2.271. View

4.
FEGLER G . The reliability of the thermodilution method for determination of the cardiac output and the blood flow in central veins. Q J Exp Physiol Cogn Med Sci. 1957; 42(3):254-66. DOI: 10.1113/expphysiol.1957.sp001261. View

5.
FEGLER G . Measurement of cardiac output in anaesthetized animals by a thermodilution method. Q J Exp Physiol Cogn Med Sci. 1954; 39(3):153-64. DOI: 10.1113/expphysiol.1954.sp001067. View