Fracture of the Polyethylene Tibial Post in Posterior Stabilized (Insall Burstein II) Total Knee Arthroplasty

Overview
Journal J Arthroplasty
Specialty Orthopedics
Date 2000 Oct 6
PMID 11021462
Citations 15
Authors
Affiliations
Soon will be listed here.
Abstract

Polyethylene wear is a well-recognized complication of total knee arthroplasty (TKA). To our knowledge, fracture of the polyethylene tibial post in a posterior stabilized TKA system has not been described as a complication or cause of failure in TKA. We report a case of a fractured polyethylene tibial post in an 85-year-old woman within 3 years of Insall Burstein II TKA, causing instability that necessitated revision surgery.

Citing Articles

Highly cross-linked polyethylene versus conventional polyethylene in primary total knee arthroplasty: comparable clinical and radiological results at a 10-year follow-up.

Giustra F, Bistolfi A, Bosco F, Fresia N, Sabatini L, Berchialla P Knee Surg Sports Traumatol Arthrosc. 2022; 31(3):1082-1088.

PMID: 36409325 PMC: 9958150. DOI: 10.1007/s00167-022-07226-6.


Causes and Clinical Outcomes of Patellar Post Impingement.

Tomite T, Saito H, Kijima H, Ishikawa N, Hatakeyama Y, Tazawa H Arthroplast Today. 2022; 14:48-52.

PMID: 35242956 PMC: 8857593. DOI: 10.1016/j.artd.2021.12.008.


Comparable results between crosslinked polyethylene and conventional ultra-high molecular weight polyethylene implanted in total knee arthroplasty: systematic review and meta-analysis of randomised clinical trials.

Bistolfi A, Giustra F, Bosco F, Faccenda C, Viotto M, Sabatini L Knee Surg Sports Traumatol Arthrosc. 2022; 30(9):3120-3130.

PMID: 35182171 PMC: 9418273. DOI: 10.1007/s00167-022-06879-7.


High rate of tibial post-fracture in highly cross-linked polyethylene compared to conventional polyethylene in posterior-stabilized total knee arthroplasty at a minimum 12-year follow-up.

Baek J, Lee S, Choi K, Nam C Knee Surg Sports Traumatol Arthrosc. 2022; 30(12):4072-4077.

PMID: 35129642 DOI: 10.1007/s00167-022-06901-y.


Fracture of the insert cone of a polyethylene liner in a failed posterior-stabilized, rotating-platform total knee arthroplasty.

Angerame M, Jennings J, Dennis D Arthroplast Today. 2018; 4(2):148-152.