Background And Purpose: Movements of the participant during blood oxygen level-dependent (BOLD) functional MR imaging cerebral activation studies are known to produce occasionally regions of false activation, especially when these movements are relatively large (>3 mm) and highly correlated with the stimulus. We investigated whether minimal (<1 mm), weakly correlated movements in a controlled functional MR imaging model could produce false activation artifacts that could potentially mimic regions of true activation in size, location, and statistical significance.

Methods: A life-size brain phantom was constructed by embedding vials of a dilute carboxylic acid solution within a gadolinium-doped gelatin mold. Imaging was performed at 1.5 T using a 2D spiral sequence (3,000/5 [TR/TE]; flip angle, 88 degrees; matrix, 64 x 64; field of view, 24 cm; section thickness, 5 mm). Controlled, in-plane, submillimeter movements of the phantom were generated using a pneumatic system and were made to correlate with a hypothetical "boxcar" stimulus over the range 0.31 < r < 0.96. Regions of false activation were sought using standard statistical methods (SPM96) that excluded phantom edges and accounted for spatial extent (regions tested at P < .05, corrected for multiple comparisons). A similar experiment was performed on a resting volunteer.

Results: The pneumatic system provided motion control with average in-plane displacements and rotations of 0.74 mm and 0.47 degrees, respectively, in the 18 data sets analyzed. No areas of false activation in the phantom were identified for poorly correlated motions (r < 0.52). Above this level, false activations occurred with increasing frequency, scaling in size and number with the degree of motion correlation. For motions with r > 0.67, areas of false activation were seen in every experiment. For a statistical threshold of P = .001, the median number of falsely activated regions was 3.5, with a mean size of 71.7 voxels (approximately 5 cc). Areas of possibly false activation of average size 72.5 voxels resulting from passive motion of the resting human participant were observed in two of four experiments.

Conclusion: Participant movements of 1 mm or less that are only modestly correlated with a blocked stimulus paradigm can produce appreciable false activation artifacts on BOLD functional MR imaging studies, even when strict image realignment methods are used to prevent them.

Citing Articles

Denoising task-correlated head motion from motor-task fMRI data with multi-echo ICA.

Reddy N, Zvolanek K, Moia S, Caballero-Gaudes C, Bright M Imaging Neurosci (Camb). 2024; 2.

PMID: 39328846 PMC: 11426116. DOI: 10.1162/imag_a_00057.


Characterizing the supraspinal sensorimotor control of walking using MRI-compatible system: a systematic review.

Hong Y, Bao D, Manor B, Zhou J J Neuroeng Rehabil. 2024; 21(1):34.

PMID: 38443983 PMC: 10913571. DOI: 10.1186/s12984-024-01323-y.


Denoising task-correlated head motion from motor-task fMRI data with multi-echo ICA.

Reddy N, Zvolanek K, Moia S, Caballero-Gaudes C, Bright M bioRxiv. 2023; .

PMID: 37503125 PMC: 10370165. DOI: 10.1101/2023.07.19.549746.


Characterizing the Action-Observation Network Through Functional Near-Infrared Spectroscopy: A Review.

Condy E, Miguel H, Millerhagen J, Harrison D, Khaksari K, Fox N Front Hum Neurosci. 2021; 15:627983.

PMID: 33679349 PMC: 7930074. DOI: 10.3389/fnhum.2021.627983.


Head motion during fMRI tasks is reduced in children and adults if participants take breaks.

Meissner T, Walbrin J, Nordt M, Koldewyn K, Weigelt S Dev Cogn Neurosci. 2020; 44:100803.

PMID: 32716852 PMC: 7284013. DOI: 10.1016/j.dcn.2020.100803.


References
1.
Bandettini P, Jesmanowicz A, Wong E, Hyde J . Processing strategies for time-course data sets in functional MRI of the human brain. Magn Reson Med. 1993; 30(2):161-73. DOI: 10.1002/mrm.1910300204. View

2.
Madsen E, Fullerton G . Prospective tissue-mimicking materials for use in NMR imaging phantoms. Magn Reson Imaging. 1982; 1(3):135-41. DOI: 10.1016/0730-725x(82)90204-1. View

3.
Bullmore E, Brammer M, Rabe-Hesketh S, Curtis V, Morris R, Williams S . Methods for diagnosis and treatment of stimulus-correlated motion in generic brain activation studies using fMRI. Hum Brain Mapp. 1999; 7(1):38-48. PMC: 6873318. View

4.
Mitchell M, KUNDEL H, Axel L, Joseph P . Agarose as a tissue equivalent phantom material for NMR imaging. Magn Reson Imaging. 1986; 4(3):263-6. DOI: 10.1016/0730-725x(86)91068-4. View

5.
Yetkin F, Haughton V, Cox R, Hyde J, Birn R, Wong E . Effect of motion outside the field of view on functional MR. AJNR Am J Neuroradiol. 1996; 17(6):1005-9. PMC: 8338627. View