Development of a Compact Proton Scanning System in Uppsala with a Moveable Second Magnet
Overview
Overview
Journal
Phys Med Biol
Publisher
IOP Publishing
Specialties
Biophysics
Nuclear Medicine
Radiology
Nuclear Medicine
Radiology
Date
2000 Jun 8
PMID
10843097
Authors
Affiliations
Affiliations
Soon will be listed here.
Abstract
A scanned proton beam yields dose distributions that in most cases are superior to passively scattered proton beams and to other external radiation treatment modalities. The present paper gives a description of the scanning system that has been developed at the Svedberg Laboratory (TSL) in Uppsala. The scanning technique and the technical design are described. The solution with a small pole gap of the magnets and a moveable second magnet results in a very compact scanning head, which can therefore be incorporated in a gantry of relatively limited size. A prototype was constructed that has been used to realize various dose distributions with a scanned beam of 180 MeV protons at TSL.