» Articles » PMID: 10796934

Endovascular Treatment of Experimental Canine Aneurysms: Feasibility with MR Imaging Guidance

Overview
Journal Radiology
Specialty Radiology
Date 2000 May 5
PMID 10796934
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate the feasibility of using magnetic resonance (MR) imaging to guide and monitor endovascular therapeutic procedures.

Materials And Methods: Endovascular therapeutic procedures were performed with MR imaging guidance in eight dogs by using a 1.5-T MR unit with echo-planar imaging capabilities. Carotid arterial aneurysms were surgically created in four dogs. The ability to depict, track, and position catheters, guide wires, and Guglielmi detachable coils was assessed. Catheters were first positioned with fluoroscopic guidance. Tracking and depiction were achieved with MR imaging by using commercially available catheters filled with a gadopentetate dimeglumine solution and a fast, two-dimensional, time-resolved, variable-rate k-space sampling technique.

Results: When either a catheter or the coaxial space between a catheter and a guide wire was filled with a solution of gadopentetate dimeglumine, catheter movement was always depicted. In the animals with aneurysms, it was possible to depict movement of a catheter into and out of the aneurysm. This was achieved by superimposing reconstructed images obtained during catheter movement onto a previously acquired MR angiogram ("road map"). Prototype Guglielmi detachable coils were successfully positioned and detached. Aneurysm obliteration was monitored with the acquisition of new road map images.

Conclusion: The results demonstrate the feasibility of using MR imaging to guide endovascular therapeutic procedures.

Citing Articles

Time-resolved imaging of contrast kinetics does not improve performance of follow-up MRA of embolized intracranial aneurysms.

Serafin Z, Strzesniewski P, Lasek W, Beuth W Med Sci Monit. 2012; 18(7):MT60-5.

PMID: 22739746 PMC: 3560766. DOI: 10.12659/msm.883199.


MR fluoroscopy in vascular and cardiac interventions (review).

Saeed M, Hetts S, English J, Wilson M Int J Cardiovasc Imaging. 2011; 28(1):117-37.

PMID: 21359519 PMC: 3275732. DOI: 10.1007/s10554-010-9774-1.


MR imaging of hydrogel filament embolic devices loaded with superparamagnetic iron oxide or gadolinium.

Killer M, Keeley E, Cruise G, Schmitt A, McCoy M Neuroradiology. 2010; 53(6):449-56.

PMID: 20625708 DOI: 10.1007/s00234-010-0744-z.


An algorithm for tracking microcatheters in fluoroscopy.

Takemura A, Hoffmann K, Suzuki M, Wang Z, Rangwala H, Harauchi H J Digit Imaging. 2007; 21(1):99-108.

PMID: 17318702 PMC: 3043820. DOI: 10.1007/s10278-007-9016-9.


Microcatheter tip enhancement in fluoroscopy: a comparison of techniques.

Takemura A, Hoffmann K, Suzuki M, Wang Z, Rangwala H, Harauchi H J Digit Imaging. 2006; 20(4):367-72.

PMID: 16946988 PMC: 3043922. DOI: 10.1007/s10278-006-0855-6.