Comparison of the MB/BacT and BACTEC MGIT 960 System for Recovery of Mycobacteria from Clinical Specimens

Overview
Date 2000 May 5
PMID 10794936
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

A total of 543 specimens were cultured in parallel with the MB/BacT and BACTEC MGIT 960 systems and on the conventional solid media. Mycobacteria were identified from 95 (17.5%) specimens, including 63 (66.3%) Mycobacterium tuberculosis and 32 (33.7%) nontuberculous mycobacteria. The recovery rates for the MB/BacT, MGIT 960, and solid media were 91.6, 87.4, and 54.7%, respectively, for all mycobacteria; the recovery rates were 93.6, 88.9, and 63.4%, respectively, for M. tuberculosis complex alone, and 87.5, 84.4, and 37.5%, respectively, for all nontuberculous mycobacteria. The mean times to detection of all mycobacteria by individual systems were 13. 9, 8.7, 31.7 days for the MB/BacT, MGIT 960 and solid media, respectively, 13.9, 9.3, 32.9 days for M. tuberculosis alone, and 14. 1, 8.1, 27.2 days for all nontuberculous mycobacteria. The contamination rates of the MB/BacT and MGIT 960 were 10.2 and 5.4%, respectively. With regard to detection times and recovery rates, both automated systems are superior to the conventional media (all p < 0.005). As compared to the MB/BacT, the MGIT 960 detected mycobacterial growth more rapidly (p < 0.001), and had a lower contamination rate (p = 0.003); however, there was no statistically significant difference in recovery rates between these two systems. These results indicate that both MGIT 960 and MB/BacT systems are rapid, sensitive, and efficient methods for the recovery of mycobacteria from clinical specimens.

Citing Articles

Probability of negative mycobacterium tuberculosis complex cultures based on time to detection of positive cultures: a multicenter evaluation of commercial-broth-based culture systems.

Tyrrell F, Budnick G, Elliott T, Gillim-Ross L, Hildred M, Mahlmeister P J Clin Microbiol. 2012; 50(10):3275-82.

PMID: 22837326 PMC: 3457445. DOI: 10.1128/JCM.01225-12.


Comparison of the conventional culture, the manual fluorescent MGIT system and the automated fluorescent MGIT 960 culture system for the detection of Mycobacterium avium ssp. avium in tissues of naturally infected hens.

Shitaye E, Beran V, Svobodova J, Moravkova M, Babak V, Pavlik I Folia Microbiol (Praha). 2009; 54(2):137-41.

PMID: 19418251 DOI: 10.1007/s12223-009-0020-y.


Meta-analysis of BACTEC MGIT 960 and BACTEC 460 TB, with or without solid media, for detection of mycobacteria.

Cruciani M, Scarparo C, Malena M, Bosco O, Serpelloni G, Mengoli C J Clin Microbiol. 2004; 42(5):2321-5.

PMID: 15131224 PMC: 404614. DOI: 10.1128/JCM.42.5.2321-2325.2004.


Susceptibility testing with the manual mycobacteria growth indicator tube (MGIT) and the MGIT 960 system provides rapid and reliable verification of multidrug-resistant tuberculosis.

Adjers-Koskela K, Katila M J Clin Microbiol. 2003; 41(3):1235-9.

PMID: 12624056 PMC: 150305. DOI: 10.1128/JCM.41.3.1235-1239.2003.


Evaluation of BacT/Alert 3D liquid culture system for recovery of mycobacteria from clinical specimens using sodium dodecyl (lauryl) sulfate-NaOH decontamination.

Carricajo A, Fonsale N, Vautrin A, Aubert G J Clin Microbiol. 2001; 39(10):3799-800.

PMID: 11574623 PMC: 88439. DOI: 10.1128/JCM.39.10.3799-3800.2001.