Temperature-dependence of the Mechanical Properties of FRP Orthodontic Wire
Overview
Affiliations
The temperature-dependence of the mechanical properties of a new esthetic orthodontic wire with fiber-reinforced plastic (FRP) structure was investigated. The new FRP wire, fabricated by a hot drawing method, is 0.5 mm in diameter and has a multiple fiber structure composed of biocompatible CaO-P2O5-SiO2-Al2O3 glass fibers of 20 microns in diameter and a polymethyl methacrylate matrix. The flexural load at a deflection of 1 mm and Young's modulus at 24, 37, and 50 degrees C under wet conditions showed similar fiber fraction dependence to those under dry conditions for a fiber fraction of 40-51%. The flexural load and Young's modulus tended to decrease slightly with increases in temperature. This tendency was larger for the lower fiber fraction. However, the difference in flexural load for a temperature difference of between 24 degrees C and 50 degrees C was at most 10 gf. This is negligibly small, and a constant orthodontic force regarding temperature change would be advantageous from a clinical point of view.
Muguruma T, Iijima M, Yuasa T, Kawaguchi K, Mizoguchi I Angle Orthod. 2016; 87(4):610-617.
PMID: 27731649 PMC: 8366697. DOI: 10.2319/022416-161.1.