Lung Cancer Screening: Minimum Tube Current Required for Helical CT

Overview
Journal Radiology
Specialty Radiology
Date 2001 Feb 7
PMID 10751484
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To investigate the minimum tube current required for helical computed tomography (CT) for lung cancer screening.

Materials And Methods: Thirty helical scans of the lung were obtained at effective tube currents of 50, 30, 20, 18, 12, 10, and 6 mAs in seven healthy volunteers. Computer-generated nodules 6 mm in diameter that showed ground-glass opacity were superimposed on the images. The image quality and detectability of nodules were evaluated subjectively by six observers. The SDs of measured CT numbers were calculated. The results were analyzed according to location in the lung.

Results: Compared with the subjective quality of images obtained at 50 mAs, the subjective quality of images obtained at 20 mAs was not significantly different. The detectability of nodules was not significantly degraded by reducing the tube current to 20 mAs in the upper zone of the lung, to 12 mAs in the middle zone, or to 18 mAs in the lower zone. The SDs at the apex and base of the lung were larger than those at other levels, and the difference became greater as the dose was reduced.

Conclusion: The minimum tube current required for screening helical CT differs for different locations in the lung. An ideal CT protocol for the lung should permit the tube current to be changed during helical scanning.

Citing Articles

Computed tomography in cystic fibrosis lung disease: a focus on radiation exposure.

Joyce S, Carey B, Moore N, Mullane D, Moore M, McEntee M Pediatr Radiol. 2021; 51(4):544-553.

PMID: 33743038 DOI: 10.1007/s00247-020-04706-0.


Environmental asbestos disease: pleural plaque volume measurement with Chest Tomography is there a correlation between pulmonary function?.

Cosgun I, Evyapan F, Karabulut N Sarcoidosis Vasc Diffuse Lung Dis. 2020; 34(4):336-342.

PMID: 32476866 PMC: 7170079. DOI: 10.36141/svdld.v34i4.5791.


Regularization Analysis and Design for Prior-Image-Based X-Ray CT Reconstruction.

Zhang H, Gang G, Dang H, Stayman J IEEE Trans Med Imaging. 2018; 37(12):2675-2686.

PMID: 29994249 PMC: 6295916. DOI: 10.1109/TMI.2018.2847250.


Computed tomography dose optimisation in cystic fibrosis: A review.

Ferris H, Twomey M, Moloney F, ONeill S, Murphy K, OConnor O World J Radiol. 2016; 8(4):331-41.

PMID: 27158420 PMC: 4840191. DOI: 10.4329/wjr.v8.i4.331.


Thoracic lymph node delineation at dose-reduced (1 mSv) dose-modulated contrast enhanced MDCT: a retrospective pilot study.

Paolini M, Wirth K, Tufman A, Reiser M, Huber R, Mueller-Lisse U Radiol Med. 2016; 121(8):644-51.

PMID: 27154525 DOI: 10.1007/s11547-016-0645-2.