Assessment of Five Portable Blood Glucose Meters, a Point-of-care Analyzer, and Color Test Strips for Measuring Blood Glucose Concentration in Dogs

Overview
Date 2000 Jan 29
PMID 10649753
Citations 15
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To compare blood glucose concentrations obtained using a point-of-care (POC) analyzer, 5 portable blood glucose meters (PBGM), and a color reagent test strip with concentrations obtained using a reference method, and to compare glucose concentrations obtained using fresh blood samples in the PBGM with concentrations obtained using blood anticoagulated with lithium heparin.

Design: Case series.

Sample Population: 110 blood samples from 34 dogs; glucose concentration of the samples ranged from 41 to 596 mg/dl.

Procedure: Logistic regression was used to compare blood glucose concentrations obtained with the various devices with reference method concentrations. Ease of use was evaluated subjectively. Percentage of times a clinical decision would have been altered if results of each of these methods had been used, rather than results of the reference method, was calculated.

Results: For 3 of the PBGM, blood glucose concentrations obtained with fresh blood were not significantly different from concentrations obtained with blood samples anticoagulated with lithium heparin. None of the devices provided results statistically equivalent to results of the reference method, but the POC analyzer was more accurate than the others. For some samples, reliance on results of the PBGM or the color test strip would have resulted in erroneous clinical decisions.

Conclusions And Clinical Relevance: Although commercially available PBGM and color test strips provided blood glucose concentrations reasonably close to those obtained with reference methods, some devices were more accurate than others. Use of results from these devices could lead to erroneous clinical decisions in some cases.

Citing Articles

Impact on result interpretation of correct and incorrect selection of veterinary glucometer canine and feline settings.

Pena L, Flatland B, Behrend E, Arzon-Pereira A, Cole J, Raz M J Vet Diagn Invest. 2023; 35(6):710-720.

PMID: 37608786 PMC: 10621561. DOI: 10.1177/10406387231195386.


Effect of site of sample collection on blood glucose concentrations measured with a portable blood glucose meter in healthy and diabetic dogs.

Jahan S, Jahan S, Jamshidi S, Linari G, Fracassi F, Akbarein H Open Vet J. 2023; 13(3):322-326.

PMID: 37026079 PMC: 10072835. DOI: 10.5455/OVJ.2023.v13.i3.8.


Evaluation of Three Human-Use Glucometers for Blood Glucose Measurement in Dogs.

Dos Santos M, Vargas A, Rosato P, Andrade C, Martins C, Petri G Vet Med Int. 2022; 2022:9112961.

PMID: 36465855 PMC: 9715321. DOI: 10.1155/2022/9112961.


Comparison of a human portable blood glucose meter and automated chemistry analyser for measurement of blood glucose concentrations in healthy dogs.

Ismail-Hamdi S, Romdane M, Romdhane S Vet Med Sci. 2021; 7(6):2185-2190.

PMID: 34352158 PMC: 8604129. DOI: 10.1002/vms3.594.


Comparison of glucose concentrations in canine whole blood, plasma, and serum measured with a veterinary point-of-care glucometer.

Suchowersky N, Carlson E, Lee H, Behrend E J Vet Diagn Invest. 2021; 33(4):695-702.

PMID: 34075827 PMC: 8229823. DOI: 10.1177/10406387211019755.