Abstract

Background: Whether the nature of haemodialysis (HD) membranes can influence the outcome of acute renal failure (ARF) remains debatable. Recent studies have suggested that dialysis with bioincompatible unsubstituted cellulosic membranes is associated with a less favourable patient outcome than dialysis with biocompatible synthetic membranes. Since we generally use a modified cellulosic membrane with substantially lower complement- and leukocyte-activating potential than cuprophane, for dialysis of patients with ARF, and because there are no data in the literature regarding the influence of modified cellulosic membranes on the outcome of patients with ARF, we compared the outcome of ARF patients dialysed either with cellulose diacetate or with a synthetic polysulfone membrane. We also investigated the potential role of permeability by comparing membranes with high-flux versus low-flux characteristics.

Methods: This prospective, randomized, single centre study included 159 patients with ARF requiring HD. Patients were stratified according to age, gender, and APACHE II score and then randomized in chronological order to one of three dialysis membranes: low-flux polysulfone, high-flux polysulfone and meltspun cellulose diacetate.

Results: Aetiologies of ARF and the prevalence of oliguria were similarly distributed among the three groups. There was no significant difference between the three groups for survival (multivariate Cox's proportional hazards model, P=0.57), time necessary to recover renal function (P=0.82), and number of dialysis sessions required before recovery (P=0.86). Multivariate analysis showed that survival was significantly influenced only by the severity of the disease state (APACHE III score, P<0.0001), but not by the nature of the dialysis membrane (P=0.57) or the presence of oliguria (P=0.24).

Conclusions: Among patients with ARF requiring HD survival and recovery time are not significantly influenced by the use of either meltspun cellulose diacetate or the more biocompatible high-flux or low-flux polysulfone. Dialysis using modified cellulose membranes is just as effective as dialysis using synthetic polysulfone membranes, but at a lower cost. In addition, the flux of the membrane did not influence patient outcome.

Citing Articles

Study on the blood flow characteristics of venous needle retention with different super-hydrophobic surface structures.

Yu Z, Liu L, Deng Y, Zhang X, Yu C Med Biol Eng Comput. 2023; 61(3):867-874.

PMID: 36627517 PMC: 9918569. DOI: 10.1007/s11517-023-02767-5.


Influence of Dialysis Membranes on Clinical Outcomes: From History to Innovation.

Chen Y, Ou S, Lin C Membranes (Basel). 2022; 12(2).

PMID: 35207074 PMC: 8876340. DOI: 10.3390/membranes12020152.


Novel Cellulose Acetate-Based Monophasic Hybrid Membranes for Improved Blood Purification Devices: Characterization under Dynamic Conditions.

Janeca A, Rodrigues F, Goncalves M, Faria M Membranes (Basel). 2021; 11(11).

PMID: 34832054 PMC: 8624022. DOI: 10.3390/membranes11110825.


Association between the recurrence period of acute kidney injury and mortality: a single-centre retrospective observational study in Japan.

Sako K, Furuichi K, Yamamura Y, Oshima M, Toyama T, Kaneko S BMJ Open. 2019; 9(6):e023259.

PMID: 31209081 PMC: 6588963. DOI: 10.1136/bmjopen-2018-023259.


The Japanese Clinical Practice Guideline for acute kidney injury 2016.

Doi K, Nishida O, Shigematsu T, Sadahiro T, Itami N, Iseki K J Intensive Care. 2018; 6:48.

PMID: 30123509 PMC: 6088399. DOI: 10.1186/s40560-018-0308-6.