Characterization of Scatter and Penetration Using Monte Carlo Simulation in 131I Imaging

Overview
Journal J Nucl Med
Specialty Nuclear Medicine
Date 2000 Jan 27
PMID 10647615
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

Unlabelled: In 131I SPECT, image quality and quantification accuracy are degraded by object scatter as well as scatter and penetration in the collimator. The characterization of energy and spatial distributions of scatter and penetration performed in this study by Monte Carlo simulation will be useful for the development and evaluation of techniques that compensate for such events in 131I imaging.

Methods: First, to test the accuracy of the Monte Carlo model, simulated and measured data were compared for both a point source and a phantom. Next, simulations to investigate scatter and penetration were performed for four geometries: point source in air, point source in a water-filled cylinder, hot sphere in a cylinder filled with nonradioactive water, and hot sphere in a cylinder filled with radioactive water. Energy spectra were separated according to order of scatter, type of interaction, and gamma-ray emission energy. A preliminary evaluation of the triple-energy window (TEW) scatter correction method was performed.

Results: The accuracy of the Monte Carlo model was verified by the good agreement between measured and simulated energy spectra and radial point spread functions. For a point source in air, simulations show that 73% of events in the photopeak window had either scattered in or penetrated the collimator, indicating the significance of collimator interactions. For a point source in a water-filled phantom, the separated energy spectra showed that a 20% photopeak window can be used to eliminate events that scatter more than two times in the phantom. For the hot sphere phantoms, it was shown that in the photopeak region the spectrum shape of penetration events is very similar to that of primary (no scatter and no penetration) events. For the hot sphere regions of interest, the percentage difference between true scatter counts and the TEW estimate of scatter counts was <12%.

Conclusion: In 131I SPECT, object scatter as well as collimator scatter and penetration are significant. The TEW method provides a reasonable correction for scatter, but the similarity between the 364-keV primary and penetration energy spectra makes it difficult to compensate for these penetration events using techniques that are based on spectral analysis.

Citing Articles

Evaluation of Improved Imaging Properties with Tungsten-Based Parallel-Hole Collimators: A Monte Carlo Study.

Islamian J, Ljungberg M World J Nucl Med. 2024; 23(2):95-102.

PMID: 38933066 PMC: 11199031. DOI: 10.1055/s-0044-1786165.


On the use of solid Ba sources as surrogate for liquid I in SPECT/CT calibration: a European multi-centre evaluation.

Tran-Gia J, Denis-Bacelar A, Ferreira K, Robinson A, Bobin C, Bonney L EJNMMI Phys. 2023; 10(1):73.

PMID: 37993667 PMC: 10665282. DOI: 10.1186/s40658-023-00582-3.


Validation of Monte Carlo I radiopharmaceutical dosimetry workflow using a 3D-printed anthropomorphic head and neck phantom.

Adam D, Grudzinski J, Bormett I, Cox B, Marsh I, Bradshaw T Med Phys. 2022; 49(8):5491-5503.

PMID: 35607296 PMC: 9388595. DOI: 10.1002/mp.15699.


Collimator and Energy Window Evaluation in Ga-67 Imaging by Monte Carlo Simulation.

Ouahman M, Errifai R, Asmi H, Bouzekraoui Y, Douama S, Bentayeb F Mol Imaging Radionucl Ther. 2020; 29(3):118-123.

PMID: 33094575 PMC: 7583743. DOI: 10.4274/mirt.galenos.2020.21549.


Optimization of Scatter Correction Method in Samarium-153 Single-photon Emission Computed Tomography using Triple-Energy Window: A Monte Carlo Simulation Study.

Asmi H, Bentayeb F, Bouzekraoui Y, Bonutti F Indian J Nucl Med. 2020; 35(3):210-215.

PMID: 33082676 PMC: 7537924. DOI: 10.4103/ijnm.IJNM_11_20.


References
1.
Jaszczak R, Greer K, Floyd Jr C, Harris C, Coleman R . Improved SPECT quantification using compensation for scattered photons. J Nucl Med. 1984; 25(8):893-900. View

2.
Macey D, Grant E, Bayouth J, Giap H, DANNA S, Sirisriro R . Improved conjugate view quantitation of I-131 by subtraction of scatter and septal penetration events with a triple energy window method. Med Phys. 1995; 22(10):1637-43. DOI: 10.1118/1.597423. View

3.
Eary J, Appelbaum F, Durack L, Brown P . Preliminary validation of the opposing view method for quantitative gamma camera imaging. Med Phys. 1989; 16(3):382-7. DOI: 10.1118/1.596347. View

4.
de Vries D, Moore S, Zimmerman R, Mueller S, Friedland B, Lanza R . Development and validation of a Monte Carlo simulation of photon transport in an Anger camera. IEEE Trans Med Imaging. 1990; 9(4):430-8. DOI: 10.1109/42.61758. View

5.
Koral K, Zasadny K, Kessler M, Luo J, Buchbinder S, Kaminski M . CT-SPECT fusion plus conjugate views for determining dosimetry in iodine-131-monoclonal antibody therapy of lymphoma patients. J Nucl Med. 1994; 35(10):1714-20. View