Sum of Segmental Bioimpedance Analysis During Ultrafiltration and Hemodialysis Reduces Sensitivity to Changes in Body Position

Overview
Journal Kidney Int
Publisher Elsevier
Specialty Nephrology
Date 1999 Aug 5
PMID 10432410
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Bioimpedance, a noninvasive technique to analyze body composition, has attracted interest in determining body hydration in hemodialysis patients. However, the so-called whole-body (wrist-to-ankle) bioimpedance analysis (WBIA) is sensitive to changes in regional fluid distribution and tends to underestimate fluid changes during ultrafiltration in hemodialysis patients. The aim of this study was to show that volume changes calculated from a new approach, that is, segmental bioimpedance analysis (SBIA), are not affected by changes in body position.

Methods: Ten male patients (age 44 +/- 8 years, target weight 70.8 +/- 10 kg) were studied during their regular hemodialysis treatment while maintaining either a sitting or a supine body position throughout the study. Extracellular volume was calculated from extracellular resistance obtained from bioimpedance data measured for a range of frequencies (5 to 500 kHz) using the Xitron BIS4000B analyzer. Wrist-to-ankle measurements were compared with segmental arm, trunk, and leg measurements.

Results: Changes in extracellular volume estimated from wrist-to-ankle measurements only reached 80 +/- 13% and 65 +/- 17% of the actual change in body mass during sitting and supine dialysis treatments, respectively. However, when segmental measurements were analyzed, the calculated change in extracellular volume was 101 +/- 6% and 100 +/- 3% of the actual change in body mass during the sitting and supine treatments, respectively.

Conclusions: SBIA properly identifies regional fluid changes and provides an appropriate measure of fluid changes caused by ultrafiltration and hemodialysis. The volume estimation based on the sum of segmental bioimpedance measurements is independent of body position, which is a prerequisite for applications in everyday practice.

Citing Articles

Development of Machine Learning-Based Web System for Estimating Pleural Effusion Using Multi-Frequency Bioelectrical Impedance Analyses.

Nose D, Matsui T, Otsuka T, Matsuda Y, Arimura T, Yasumoto K J Cardiovasc Dev Dis. 2023; 10(7).

PMID: 37504547 PMC: 10380905. DOI: 10.3390/jcdd10070291.


Assessment of Body Composition in Athletes: A Narrative Review of Available Methods with Special Reference to Quantitative and Qualitative Bioimpedance Analysis.

Campa F, Toselli S, Mazzilli M, Gobbo L, Coratella G Nutrients. 2021; 13(5).

PMID: 34065984 PMC: 8150618. DOI: 10.3390/nu13051620.


Knee-to-knee Bioimpedance Measurements to Monitor Changes in Extracellular Fluid in Haemodynamic-unstable Patients During Dialysis.

Ismail A, Schlieper G, Walter M, Floege J, Leonhardt S J Electr Bioimpedance. 2021; 10(1):55-62.

PMID: 33584883 PMC: 7531213. DOI: 10.2478/joeb-2019-0008.


Monitoring transcellular fluid shifts during episodes of intradialytic hypotension using bioimpedance spectroscopy.

Ismail A, Gross T, Schlieper G, Walter M, Eitner F, Floege J Clin Kidney J. 2021; 14(1):149-155.

PMID: 33564413 PMC: 7857820. DOI: 10.1093/ckj/sfz123.


Prediction of hemodialysis vascular access failure using segmental bioimpedance analysis parameters.

Kim H, Seo H, Kim J, Kim M Int Urol Nephrol. 2018; 50(5):947-953.

PMID: 29476433 DOI: 10.1007/s11255-018-1827-8.