Inappropriate Implantable Defibrillator Discharge Caused by a Retained Pacemaker Lead Fragment

Overview
Publisher Springer
Date 1999 Jul 1
PMID 10387144
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Inappropriate discharge is still a major issue of implantable cardioverter defibrillator therapy. The diagnostic options of modern devices facilitate classification of the underlying abnormality.

Methods And Results: A 65-year-old woman with depressed left ventricular performance received spurious shocks from an ICD, implanted for ventricular tachycardia. A lead fragment of an explanted VVI pacemaker system could be identified as cause of erroneous ventricular fibrillation detection by the ICD. The electrical noise caused by interaction between the lead remnant and the ICD lead was detectable even on the surface ECG.

Conclusion: Based on our findings, removal of fragmented lead material should be considered prior to ICD implantation, to avoid potentially adverse and harmful interactions with ICD systems.

Citing Articles

Incessant ventricular tachycardia, refractory to catheter ablation, in an ICD patient terminated by ICD lead extraction: a case report.

Lenz C, Dietze T, Moller M, Schobel W, Wicke J, Kellner H Clin Res Cardiol. 2009; 98(12):803-5.

PMID: 19714288 DOI: 10.1007/s00392-009-0067-4.

References
1.
Swerdlow C, Chen P, Kass R, ALLARD J, Peter C . Discrimination of ventricular tachycardia from sinus tachycardia and atrial fibrillation in a tiered-therapy cardioverter-defibrillator. J Am Coll Cardiol. 1994; 23(6):1342-55. DOI: 10.1016/0735-1097(94)90376-x. View

2.
Brooks R, Garan H, McGovern B, Ruskin J . Implantation of transvenous nonthoracotomy cardioverter-defibrillator systems in patients with permanent endocardial pacemakers. Am Heart J. 1995; 129(1):45-53. DOI: 10.1016/0002-8703(95)90041-1. View

3.
Jung W, Luderitz B . Implantation of an arrhythmia management system for ventricular and supraventricular tachyarrhythmias. Lancet. 1997; 349(9055):853-4. DOI: 10.1016/S0140-6736(05)61758-8. View

4.
Wolpert C, Jung W, Spehl S, Korte T, Luderitz B . Case report: inappropriate discharge of an implantable cardioverter-defibrillator caused by the combined count criterion. J Interv Card Electrophysiol. 1998; 2(1):53-6. DOI: 10.1023/a:1009716907852. View

5.
Grimm W, Flores B, Marchlinski F . Electrocardiographically documented unnecessary, spontaneous shocks in 241 patients with implantable cardioverter defibrillators. Pacing Clin Electrophysiol. 1992; 15(11 Pt 1):1667-73. DOI: 10.1111/j.1540-8159.1992.tb02953.x. View