Anatomic M-mode Echocardiography: A New Approach to Assess Regional Myocardial Function--A Comparative in Vivo and in Vitro Study of Both Fundamental and Second Harmonic Imaging Modes

Overview
Specialty Radiology
Date 1999 May 8
PMID 10231615
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To evaluate the accuracy of anatomic M-mode echocardiography (AMM).

Methods: Eight phantoms were rotated on a device at different insonation depths (IDs) in a water beaker. They were insonated with different transducer frequencies in fundamental imaging (FI) and second harmonic imaging (SHI), and the diameters were assessed with conventional M-mode echocardiography (CMM) and AMM with the applied angle correction (AC) after rotation. In addition, left ventricular wall dimensions were measured with CMM and AMM in FI and SHI in 10 volunteers.

Results: AC had the greatest effect on the measurement error in AMM followed by ID (AC: R2 = 0. 295, ID: R2 = 0.268; P <.0001). SHI improved the accuracy, and a difference no longer existed between CMM and AMM with an AC up to 60 degrees. In vivo the limit of agreement between AMM and CMM was -1.7 to +1.8 mm in SHI.

Conclusion: Within its limitations (AC < 60 degrees; ID < 20 cm), AMM could be a robust tool in clinical practice.

Citing Articles

Four-dimensional computed tomography of the left ventricle, Part II: Estimation of mechanical activation times.

Manohar A, Pack J, Schluchter A, McVeigh E Med Phys. 2022; 49(4):2309-2323.

PMID: 35192200 PMC: 9007845. DOI: 10.1002/mp.15550.


Improving ultrasonic measurement of diaphragmatic excursion after cardiac surgery using the anatomical M-mode: a randomized crossover study.

Pasero D, Koeltz A, Placido R, Lima M, Haun O, Rienzo M Intensive Care Med. 2015; 41(4):650-6.

PMID: 25573500 DOI: 10.1007/s00134-014-3625-9.


Measurements of left ventricular myocardial longitudinal systolic displacement using spectral and colour tissue Doppler: time for a reassessment?.

Manouras A, Shahgaldi K, Winter R, Brodin L, Nowak J Cardiovasc Ultrasound. 2009; 7:12.

PMID: 19292894 PMC: 2661320. DOI: 10.1186/1476-7120-7-12.


Relevance of tissue Doppler in the quantification of stress echocardiography for the detection of myocardial ischemia in clinical practice.

Sicari R Cardiovasc Ultrasound. 2005; 3:2.

PMID: 15679889 PMC: 548514. DOI: 10.1186/1476-7120-3-2.


Techniques for comprehensive two dimensional echocardiographic assessment of left ventricular systolic function.

Marwick T Heart. 2003; 89 Suppl 3:iii2-8.

PMID: 14594869 PMC: 1876300. DOI: 10.1136/heart.89.suppl_3.iii2.