» Articles » PMID: 10050943

On Spatially Selective RF Excitation and Its Analogy with Spiral MR Image Acquisition

Overview
Journal MAGMA
Publisher Springer
Date 1999 Mar 2
PMID 10050943
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

The basic principles of the design of spatially selective RF pulses are described, and their analogy with MR image acquisition and reconstruction is shown. The paper focuses on RF-pulse design and imaging schemes in which spiral k-space trajectories are used. The sensitivity of RF excitation to gradient-system imperfections and to spatially varying off-resonance are analyzed, and suitable measures of correction are discussed. The spatial resolution obtainable with selective RF pulses and the consequences of the linearity of the pulse-design problem are examined. Phantom experiments showing the performance of multidimensional spatially selective RF pulses further illustrate the analogy with MR image acquisition.

Citing Articles

Multidimensional RF pulse design with consideration of concomitant field effects.

Zhao Z, Lee N, Nayak K Magn Reson Med. 2024; 93(2):718-729.

PMID: 39365913 PMC: 11604853. DOI: 10.1002/mrm.30311.


Parallel transmit optimized 3D composite adiabatic spectral-spatial pulse for spectroscopy.

He X, Auerbach E, Garwood M, Kobayashi N, Wu X, Metzger G Magn Reson Med. 2021; 86(1):17-32.

PMID: 33497006 PMC: 8545499. DOI: 10.1002/mrm.28682.


Improved respiratory self-navigation for 3D radial acquisitions through the use of a pencil-beam 2D-T -prep for free-breathing, whole-heart coronary MRA.

Coristine A, Chaptinel J, Ginami G, Bonanno G, Coppo S, van Heeswijk R Magn Reson Med. 2017; 79(3):1293-1303.

PMID: 28568961 PMC: 5931377. DOI: 10.1002/mrm.26764.


Correction of parallel transmission using concurrent RF and gradient field monitoring.

Cavusoglu M, Dietrich B, Brunner D, Weiger M, Pruessmann K MAGMA. 2017; 30(5):473-488.

PMID: 28444504 DOI: 10.1007/s10334-017-0620-5.


Application of the limited-memory quasi-Newton algorithm for multi-dimensional, large flip-angle RF pulses at 7T.

Vinding M, Brenner D, Tse D, Vellmer S, Vosegaard T, Suter D MAGMA. 2016; 30(1):29-39.

PMID: 27485854 DOI: 10.1007/s10334-016-0580-1.


References
1.
Twieg D . The k-trajectory formulation of the NMR imaging process with applications in analysis and synthesis of imaging methods. Med Phys. 1983; 10(5):610-21. DOI: 10.1118/1.595331. View

2.
Aldefeld B, Bornert P . Effects of gradient anisotropy in MRI. Magn Reson Med. 1998; 39(4):606-14. DOI: 10.1002/mrm.1910390414. View

3.
Liu Y, Riederer S, Rossman P, Grimm R, Debbins J, Ehman R . A monitoring, feedback, and triggering system for reproducible breath-hold MR imaging. Magn Reson Med. 1993; 30(4):507-11. DOI: 10.1002/mrm.1910300416. View

4.
Morrell G, Macovski A . Three-dimensional spectral-spatial excitation. Magn Reson Med. 1997; 37(3):378-86. DOI: 10.1002/mrm.1910370314. View

5.
Bornert P, Schaffter T . Curved slice imaging. Magn Reson Med. 1996; 36(6):932-9. DOI: 10.1002/mrm.1910360616. View